Feature
(1) Fine low-power LCD display is adopted, and instantaneous flow, umulative flow, frequency and other information can be local display.
(2) The Sensor adopts 4 pieces of piezoelectric ceramic technology to effectively eliminate the impact of external vibration on the measurement.
(3) No moving parts, no wear, no mechanical maintenance, less leakage points, long service life.
(4) The sensor and signal converter are interchangeable.
(5) The split converter can be used, and the cable length is up to 10 meters.
(6) Output 4-20mA current signal/pulse signal and ModbusRS485 differential signal.
(7) Explosion-proof: Exia II CT6 for Intrinsically safe
Exd II CT6 for Flameproof
(8) HART Protocol or ModbusRS485 interface.
(9) Can Receive temperature sensor PT1000 signal and pressure sensor signal, and output 4-20mA current signal for P/T compensation volume.
(10) High measurement accuracy: 0.5%.
(11) Requirements for straight pipe section: inlet 3D/outlet 1D.
Technical parameters
Meter Size (mm)
DN15 ~ DN400
Applicable media
Gas, liquid, steam
Nominal pressure (MPa)
1.6、2.5、4.0、6.3
other pressure with agreements
Medium temperature
Standard type (- 40~+150) ℃
Medium temperature type (- 40~+250) ℃
High temperature type (- 40~+350) ℃
Body material
(304/316/316L) stainless steel,
other materials under agreement
Accuracy
0.5%,
1.0%
Repeatability
≤ ± 0.2% (measured value)
Range ratio
1:20
turn
change
implement
form
Integrated type
Power supply
12VDC/24VDC or battery power supply
display
lcd
communication
HART protocol
Modbus RS485
output
l Pulse
3-wire
l 4-20mA
2-wire
Explosion-proof
Intrinsically safe
Exia Ⅱ CT6
flame-proof Exd Ⅱ CT6
Protection grade
IP65,IP67,IP68
Electrical interface
M20 × 1.5,1/2NPT
environment
condition
temperature
(-30 ~ +60)℃
humidity
5% ~ 90%
atmospheric pressure
(86 ~ 106)kPa